Report No 088213R-ITCEP07V04
Page: 103 of 131
Product Notebook
Test Item Electrical fast transient/burst
Test Mode Mode 2
Date of Test 2008/08/27 Test Site No.6 Shielded Room
Inject
Line
Polarity
Voltage
kV
Inject
Time
(Second)
Inject
Method
Required
Criteria
Complied
to
Criteria
Result
L-N-PE
±
1kV 60 Direct B B
PASS
LAN
±
0.5kV 60 Clamp B A
PASS
Tel ec om
±
0.5kV 60 Clamp B B
PASS
Note:
The testing performed is from lowest level up to the highest level as required by
standard, but only highest level is shown on the report.
Meet criteria A : Operate as intended during and after the test
Meet criteria B : Operate as intended after the test
Meet criteria C : Loss/Error of function
Additional Information
EUT stopped operation and could / could not be reset by operator at kV of
Line
.
No false alarms or other malfunctions were observed during or after the test.