MSI MS-9652 Manual do Utilizador Página 87

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 126
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 86
Report No 094061R-ITCEP07V06
Page: 84 of 123
8.4. Test Procedure
Direct application of discharges to the EUT:
Contact discharge was applied only to conductive surfaces of the EUT.
Air discharges were applied only to non-conductive surfaces of the EUT.
During the test, it was performed with single discharges. For the single discharge
time between successive single discharges will be keep longer 1 second. It was at
least ten single discharges with positive and negative at the same selected point.
The selected point, which was performed with electrostatic discharge, was marked
on the red label of the EUT.
Indirect application of discharges to the EUT:
Vertical Coupling Plane (VCP):
The coupling plane, of dimensions 0.5m x 0.5m, is placed parallel to, and positioned
at a distance 0.1m from, the EUT, with the Discharge Electrode touching the
coupling plane.
The four faces of the EUT will be performed with electrostatic discharge. It was at
least ten single discharges with positive and negative at the same selected point.
Horizontal Coupling Plane (HCP):
The coupling plane is placed under to the EUT. The generator shall be positioned
vertically at a distance of 0.1m from the EUT, with the Discharge Electrode touching
the coupling plane.
The four faces of the EUT will be performed with electrostatic discharge. It was at
least ten single discharges with positive and negative at the same selected point.
8.5. Deviation from Test Standard
No deviation.
Vista de página 86
1 2 ... 82 83 84 85 86 87 88 89 90 91 92 ... 125 126

Comentários a estes Manuais

Sem comentários