MSI MS-9652 Manual do Utilizador Página 92

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 126
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 91
Report No 094061R-ITCEP07V06
Page: 89 of 123
9.4. Test Procedure
The EUT and load, which are placed on a table that is 0.8 meter above ground, are
placed with one coincident with the calibration plane such that the distance from
antenna to the EUT was 3 meters.
Both horizontal and vertical polarization of the antenna and four sides of the EUT are set
on measurement.
In order to judge the EUT performance, a CCD camera is used to monitor EUT screen.
All the scanning conditions are as follows:
Condition of Test Remarks
1. Field Strength 3 V/m Level 2
2. Radiated Signal AM 80% Modulated with 1kHz
3. Scanning Frequency 80MHz - 1000MHz
4 Dwell Time 3 Seconds
5.
Frequency step size
f
:
1%
6. The rate of Swept of Frequency 1.5 x 10
-3
decades/s
9.5. Deviation from Test Standard
No deviation.
Vista de página 91
1 2 ... 87 88 89 90 91 92 93 94 95 96 97 ... 125 126

Comentários a estes Manuais

Sem comentários